* * * FOR ALL
TASK 72-00-31-800-801
1 . Eddy Current Inspection.
A.This procedure gives instructions to do a manual eddy current inspection of the coated seal teeth on the high pressure compressor rotor (HPCR) CDP Seal for cracking at the tip of the seal teeth. Refer to Figure 201.
B.This procedure can be used module level, on coated seal teeth.
C.Personnel that will do this procedure must do the requirements that follow:
(1)Personnel that do this inspection must be certified by one of the certifications that follow: NAS-410, American Society of Nondestructive Testing (ASNT-TC-1A), Air Transport Association Specification No. 105 (ATA 105), COSAC, or one of the equivalent certification document acknowledged by the local regulatory agencies.
(2)Make sure that the personnel that do this inspection receive practical training to use this procedure and must demonstrate proficiency in the calibration, inspection, and evaluation routines before accept/reject authority is delegated.
NOTE:
Training personnel on this inspection does not imply that they have completed the requirements for an inspector certification or agree with the appropriate certification document.
NOTE:
Personnel that has received the practical training on other, similar inspection (i.e. CF6-80A/-C2/-E Eddy Current Inspection of the seal teeth on the HPT thermal shield or HPT rotating interstage seal), can be considered as qualified according to TASK 72-00-31-800-801 (paragraph 1.C.(2)) of this document.
2 . Tools, Equipment, and Materials.
NOTE:
You can use equivalent alternatives for tools, equipment, and consumable materials. However, the repair source must demonstrate and document equivalency.
A.Tools and Equipment.
(1)Special Tools. None.
(2)Standard Tools and Equipment.
 
Description
Manufacturer
 
GE Aircraft Engines Inspection Kit GE FQAP-587F
GE Field Support Technology ndttooling@ge.com
 
Calibration Standard PN SGP-112 (part of inspection kit GE FQAP-587F)
--
 
Calibration Standard Holder PN SGP-113 (part of inspection kit GE FQAP-587F)
--
 
Eddy Current Probe PN US-2056 (part of inspection kit GE FQAP-587F)
--
 
Eddy Current Probe Holder (Spring Included) PN JO57138 (part of inspection kit GE FQAP-587F)
--
 
Probe Cable PN 94161 (part of inspection kit GE FQAP-587F)
--
 
Teflon Tape PN 5490 (part of inspection kit GE FQAP-587F)
--
 
Black Action Marker (size 33 fine) Q404 (part of inspection kit GE FQAP-587F)
--
 
Storm Case PN iM2100 (part of inspection kit GE FQAP-587F)
--
 
Label (part of inspection kit GE FQAP-587F)
--
NOTE:
Customers that already have the inspection kit P/N: GE-FQAP-587, do not need to order kit P/N: GE-FQAP-587F.
(3)Locally Manufactured Tools. None.
B.Consumable Materials.
 
Code No.
Description
 
C04-003
Solvent, General (acetone)
 
C04-035
Solvent, General (isopropyl alcohol)
 
--
Tape, Teflon
C.Referenced Procedures. None.
D.Expendable Parts. None.
E.SPD Information. None.
F.Special Solutions. None.
G.Test Specimens. None.
3 . Dimensional Information.
None.
4 . Setup Information.
Subtask 72-00-31-350-001
CAUTION:
SEAL TEETH ARE EASILY DAMAGED. USE EXTREME CARE IN HANDLING THE PART.
NOTE:
Parts with blends in the seal teeth, damage or conditions that prevent the probe from moving across the seal teeth cannot be eddy current inspected and must be inspected with FPI by the engine manual.
NOTE:
Make sure that the seal teeth are in the coated condition prior to eddy current inspection.
A.Make sure that the part number and the serial number of the CDP seal. Refer to Figure 201 and do as follows:
Subtask 72-00-31-350-002
(1)Put a reference mark on the aft side of the part close to the seal teeth that is radially in line with the slash symbol (/) in the serial number identification (S/N).
NOTE:
If S/N is not visible, place an arbitrary mark for the start position.
NOTE:
This will be the start and stop location for each seal tooth scan.
Subtask 72-00-31-220-081
WARNING:
REFER TO THE PRODUCT LABEL AND THE MANUFACTURER'S (MATERIAL) SAFETY DATA SHEET (SDS) FOR INSTRUCTIONS ON THE HAZARDS, STORAGE, SAFE HANDLING AND PROPER USE OF THESE PRODUCTS.
(2)Perform a visual inspection of the seal teeth and mark on the part the location of any nicks, dents or other anomalies, which may prevent the inspection or cause an indication during inspection. Any damage or condition that prevents the probe from moving across the tooth must be inspected with FPI by the engine manual. If necessary, lightly clean the areas using an approved solvent.
5 . Procedure.
Subtask 72-00-31-250-001
A.Do the initial equipment setup as follows:
(1)Turn the eddy current instrument on.
(2)Let the instrument warm up. Refer to manufacturer's recommendations or at least 10 minutes before the inspection.
(3)Adjust the instrument to the initial settings listed in the table that follows:
 
STAVELEY - NORTEC 19e & 19eII
 
CONTROL DESCRIPTION
INITIAL SETTING
 
MAIN
 
FREQUENCY
500 KHZ
 
ANGLE
350
 
H-DB
51 DB
 
V-DB
51 DB
 
PROBE DRIVE
HIGH
 
DISPLAY
F1
 
SWEEP
OFF
 
V-POS
50%
 
H-POS
50%
 
SCREEN
 
PERSIST
OFF
 
DISP ERASE
AS NEEDED
 
FILTER
 
LPF
100 HZ
 
HPF
0
 
STAVELEY - NORTEC 24
 
CONTROL DESCRIPTION
INITIAL SETTING
 
FREQUENCY
500 KHZ
 
GAIN
 
H-DB
50 DB
 
V-DB
50 DB
 
PROBE DRIVE
HIGH
 
ANGLE
170
 
DISPLAY
 
SWEEP
OFF
 
V-POS
50%
 
H-POS
50%
 
SCREEN
 
PERSIST
OFF
 
SWP ERS
OFF
 
DISP ERS
AS NEEDED
 
FILTER
 
LP FILT
100 HZ
 
HP FILT
0
 
HPF POLE
2-POLE
 
STAVELEY - NORTEC 2000 Series
 
CONTROL DESCRIPTION
INITIAL SETTING
 
MAIN
 
FREQUENCY
500 KHZ
 
ANGLE
170
 
H-DB
50.0 DB
 
V-DB
50.0 DB
 
PROBE DRIVE
HIGH
 
CONT NULL
OFF
 
DISPLAY
 
SWEEP
OFF
 
V-POS
50%
 
H-POS
50%
 
SCREEN
 
PERSIST
OFF
 
DISP ERASE
AS NEEDED
 
FILTER
 
LPF
100 HZ
 
HPF
0
 
HPF POLE
2-POLE
 
NORTEC 500 Series
 
ADAPTOR P/N: 9230392
 
CONTROL DESCRIPTION
INITIAL SETTING
 
FREQ 1 SETUP
 
FREQ
500 KHz
 
ANGLE
170
 
GAIN - H
50
 
GAIN - V
50
 
FILTER SETUP
 
HiPASS
OFF
 
LowPASS
100
 
CONT NULL
OFF
 
PRE AMP
OFF
 
BALANCE
OFF
 
DISPLAY SCREEN SETUP
 
H POS
50.00%
 
V POS
50.00%
 
DOT/BOX
DOT
 
ERASE SETUP
 
PERSIST
OFF
 
DISP ERS
AS NEEDED
 
SWP ERS
OFF
 
GRATICUL
ON
 
MAIN DSP
F1
 
SCAN
 
SWP MODE
OFF
 
SETUP
 
INST. SETUP
 
PROBE DRIVE
HIGH
 
ANGSTEP
1
 
FRQ MODE
SINGLE
 
HMIX GN
0 dB
 
VMIX GN
0 dB
 
INST. PREFS
 
LOCK
OFF
 
LANG
AS NEEDED
 
SOFTKEY
RIGHT
 
AUTO ERASE
ON
 
UNIWEST 454 Series; Pro and Premium Series
 
CONTROL DESCRIPTION
INITIAL SETTING
 
FREQ 1
500 KHz
 
F1 Gain
27
 
F1 Rot
290
 
FREQ 2
Off
 
DRIVE
High
 
LP FLT
100 Hz
 
HP FLT
Off
 
X Sens
.5 v/Div
 
Y Sens
=X
 
X-Null Point
0.0 Div
 
Y-Null Point
0.0 Div
 
AUTO CLEAR
AS NEEDED
 
ROHMANN ELOTEST B1 V3 Series
 
Adapter P/N: B1-NOB 8H
 
CONTROL DESCRIPTION
INITIAL SETTING
 
FREQ
500 KHZ
 
BW
HF
 
BW
AUTO
 
GAIN
24 DB
 
PREAMP
6 DB
 
Y AXIS SPREAD
0 DB
 
X AXIS SPREAD
0 DB
 
PHASE
290
 
DISP:
Y/X
 
FILTER U
 
HIGHPASS
STATIC
 
LOWPASS
100 HZ
 
DOT POSITION X and Y
0
 
ROHMANN ELOTEST B1 V4 Series
 
Adapter P/N: B1-NOB 8H
 
CONTROL DESCRIPTION
INITIAL SETTING
 
FREQ
500 KHZ
 
BWL
ON
 
GAIN
28/28 DB
 
PREAMP
10 DB
 
Y AXIS SPREAD
0 DB
 
X AXIS SPREAD
0 DB
 
PHASE
240
 
DISP:
Y/X
 
FILTER
 
HIGHPASS
STATIC
 
LOWPASS
100 HZ
 
DOT POSITION X & Y
0
 
ROHMANN ELOTEST B300 Series
 
Adapter P/N: B1-NOB 8H
 
CONTROL DESCRIPTION
INITIAL SETTING
 
PARAMETER SETUP
 
FREQUENCY
500 KHZ
 
AMPLITUDE
50%
 
PREAMPLIFIER
10.0 DB
 
MAIN AMPLIFIER
16.0 DB
 
SPREAD Y
0.0 DB
 
TOTAL GAIN Y/X
26.0 DB
 
PHASE
40
 
FILTER
 
LOPASS
100 HZ
 
HIPASS
STATIC
 
DOT POSITION Y
0.00%
 
DOT POSITION X
0.00%
 
PERSISTANCE
AS NEEDED
 
PRB/CHAN SETUP
 
F2
CHA1
 
INPUT A
 
PROBE TYPE
NOT USED
 
ATTENUATOR
OFF
 
HOCKING - PHASEC 2200
 
Probe Adapter P/N: 640-015-007
 
CONTROL DESCRIPTION
INITIAL SETTING
 
MODE
DIFF 1CH
 
DISPLAY
XY
 
VIEW
CH 1
 
PERSIST
AS NEEDED
 
CH 1 FREQ
500 KHZ
 
CH 1 PHASE
226
 
CH 1 GAIN
30.0 DB
 
CH 1 X:Y
0.0DB
 
HI-PASS
DC
 
LO-PASS
100 HZ
 
INP. GAIN
+20DB
 
X & Y POS
0
 
SUBMENU
 
GRATICULE
RECT.
 
PROBE
STANDARD
 
DRIVE
0 DB 2.0 V
 
HOCKING PHASEC 2s
 
Probe Adapter P/N: 640-015-007
 
CONTROL DESCRIPTION
INITIAL SETTING
 
GAIN DB
28.0/28.0
 
X DB
28
 
Y DB
28
 
PHASE
70
 
FREQUENCY
500 KHZ
 
HIPASS HZ
DC
 
LOPASS HZ
100
 
BAND LOCK
OFF
 
PROBE
BRIDGE
 
LOAD μH
22
 
INPUT GAIN
LOW
 
DISPLAY
SPOT
 
SWEEP
N/A
 
PERSIST
AS NEEDED
 
GRATICULE
GRID 2
 
SPOT X
50%
 
SPOT Y
50%
 
HOCKING PHASEC 2d
 
Probe Adapter P/N: 640-015-007
 
CONTROL DESCRIPTION
INITIAL SETTING
 
GAIN XY1
28.0/28.0
 
PHASE1
260
 
FREQ1
500 KHZ
 
FILTER1
BP/RATIO
 
HP/LP1
DC/100
 
PROBE
BRIDGE
 
DISPLAY
SPOT
 
GRATICULE
GRID 2
 
SPOT XY
90/0
 
SPOT INFO
NONE
 
PERSIST
AS NEEDED
 
SWEEP
N/A
 
LOAD μH
22
Subtask 72-00-31-820-001
B.Do the calibration of the system as follows:
(1)Insert probe P/N: US-2056 into probe holder P/N: JO57138 for the following.
(a)Make sure the spring is placed over the cable connection side of the probe and is resting on the large diameter of the probe body. Refer to Figure 204.
(b)Grasp the clamp on the probe holder by the two notched areas and slide it back, (open), until it stops. Refer to Figure 205.
(c)Insert the probe, coil end first into the probe holder. Refer to Figure 206.
(d)Make sure the flat sides of the probe body are parallel with the clamp slot and the spring is completely to the bottom of the probe holder counter bored hole. The clamp must glide easily over the top of the spring when closing to keep the probe locked into the holder.
(e)Connect the inspection probe, P/N: US-2056, to the eddy current instrument with the probe cable, P/N: 94161.
(2)Apply Teflon tape to the probe as follows:
(a)Make sure the tape is correctly attached to the bottom of the groove in the probe tip. Refer to Figure 206.
(b)An acceptable method to attach the tape into the groove is with the use of the calibration standard to push the tape into the bottom of the probe tip groove and swiping the probe along the standard.
(c)Make sure that the tape does not inhibit movement of the probe inside the probe holder.
CAUTION:
DO NOT PUT TENSION ON THE PROBE CABLE. TENSION WILL CAUSE THE PROBE TO PULL AWAY FROM THE STANDARD SURFACE AND CAN CAUSE A LIFT-OFF EFFECT AND CHANGE IN CALIBRATION SENSITIVITY.
NOTE:
Deleted.
(3)Put the eddy current probe and probe holder on the calibration standard until the inspection coil is away from the electro discharge machining (EDM) notch.
(4)Make sure the probe holder is seated flat on the calibration standard holder and the probe spring compresses up when set on the calibration standard. Refer to Figure 202, Figure 203, and Figure 206.
(5)Null the eddy current instrument.
(6)Slowly scan the probe across the EDM notch until the notch is detected and observe the response. Scan speed must be kept relatively slow, approximately 3.0 inches (76 mm) per second or less.
NOTE:
While you scan across the EDM notch in different directions, a small difference in notch response can be observed. The highest amplitude response (regardless of scan direction) can be used to set the calibration sensitivity.
(7)While scanning back and forth across the EDM notch, adjust the instrument phase and gain control until the EDM notch response is vertical and 6 major divisions peak to peak. Refer to Figure 208.
(8)Null the eddy current instrument again after you do instrument adjustments. Make sure that the probe coil is away from the EDM notch.
(9)Scan the EDM notch again to make sure that the notch response is still vertical and at 6 major divisions peak to peak.
(10)Record all instrument settings and requested information on the appropriate inspection data sheet. Refer to Figure 210.
Subtask 72-00-31-250-002
C.Do the hardware inspection as follows:
NOTE:
Parts with blends in the seal teeth, damage or conditions that prevent the probe from moving across the seal teeth cannot be eddy current inspected and must be inspected with FPI by the engine manual.
NOTE:
Make sure that the seal teeth are in the coated condition prior to do the eddy current inspection.
(1)Calibrate the equipment. Refer to Subtask 72-00-31-820-001 (paragraph 5.B).
Subtask 72-00-31-250-003
CAUTION:
DO NOT PUT TENSION ON THE PROBE CABLE. TENSION WILL CAUSE THE PROBE TO PULL AWAY FROM THE SEAL TOOTH SURFACE CAUSING A LIFT OFF EFFECT AND CHANGE IN INSPECTION SENSITIVITY.
(2)Start at the previously marked reference point (S/N). Refer to Subtask 72-00-31-350-002 (paragraph 4.A.(1)).
(3)Put the eddy current probe on the number one seal tooth. Refer to Figure 201.
(4)Maintain the probe at an angle similar to the angle of the seal teeth. The bubble level is not used for this inspection. Refer to Figure 207.
(5)When you put the coil over the seal tooth, make sure the spring applies pressure to the probe and the seal tooth is riding in the groove of both the probe and the probe holder wheels. Refer to Figure 207.
(6)Try to keep the wheels of the probe holder on the seal teeth so the probe is not restricted while performing the inspection around the seal teeth.
(7)With the probe properly seated over the seal tooth, null the eddy current instrument.
(8)Scan the circumference of the part overlapping the start and stop position to make sure a 360 degree coverage of the entire seal tooth. Scan speed must be kept relatively slow, approximately 3.0 inches (76 mm) per second or less.
NOTE:
The scans do not have to be continuous for 360 degrees. The seal tooth can be divided into sections and each section inspected. If this is done, make sure that all scans overlap.
(9)Monitor the instrument display for any indication equal to or exceeding 3.0 major divisions, regardless of scan direction.
(10)For any indications equal to or exceeding 3.0 major divisions, perform an evaluation as described in the evaluation section of this document. Record the indication amplitude on the inspection data sheet. Refer to Figure 210.
Subtask 72-00-31-250-004
(11)It is recommended that the tape be replaced after every tooth is inspected or as needed. A calibration check must be performed before and after every tape change to make sure that the calibration sensitivity has not changed. Refer to Subtask 72-00-31-210-001 (paragraph 5.D.) for a calibration check.
Subtask 72-00-31-250-005
(12)Do Subtask 72-00-31-250-003 (paragraph 5.C.(2)) thru Subtask 72-00-31-250-004 (paragraph 5.C.(11)) again for the remaining seal teeth.
Subtask 72-00-31-210-001
D.Do a calibration check as follows:
(1)Do a calibration check, if you are in the conditions that follow:
(a)After each inspection.
(b)The component is changed.
(c)The operator is changed.
(d)After loss of power.
(e)When the operator suspects a change in the system.
(f)Before and after every tape change.
(g)At least once every hour.
(2)Put the eddy current probe and probe holder on the calibration standard until the inspection coil is away from the EDM notch.
(3)Make sure that the probe holder is setting flat on the calibration standard holder and that the probe spring is compressed when set on the standard.
(4)Null the eddy current instrument.
(5)Slowly scan the probe across the EDM notch of the calibration standard and observe the response.
NOTE:
While you scan across the EDM notch in different directions, a small difference in notch response can be observed. The highest amplitude response (regardless of scan direction) can be used to check the calibration sensitivity.
(6)If the calibration shows to be in a half division full screen height (FSH) (5.5 to 6.5) of the initial calibration again, the test was acceptable.
(7)If the notch response is less than 5.5 divisions, the system must be calibrated again and all hardware inspected since the last acceptable calibration must be inspected again.
(8)If the notch response is greater than 6.5 divisions, the system must be calibrated again prior to inspection of additional hardware and any rejected hardware must be inspected again.
Subtask 72-00-31-210-002
E.Do an evaluation of the indications as follows:
(1)Indications that are equal or more than three major divisions must be evaluated as follows:
(a)Make sure that the calibration notch response is still at 6 major divisions in amplitude.
(b)Do a visual inspection of the seal tooth for any blends, nicks, dents, or other anomalies. Record the results on Figure 210.
WARNING:
REFER TO THE PRODUCT LABEL AND THE MANUFACTURER'S (MATERIAL) SAFETY DATA SHEET (SDS) FOR INSTRUCTIONS ON THE HAZARDS, STORAGE, SAFE HANDLING AND PROPER USE OF THESE PRODUCTS.
CAUTION:
DO NOT USE SCOTCH-BRITE. DAMAGE TO THE PART CAN OCCUR.
(c)If necessary, lightly clean the surface in question with C04-003   acetone or C04-035  isopropyl alcohol.
(d)If the suspect signal remains but is reduced in amplitude repeat the cleaning process as required to reduce the amplitude to an acceptable level.
(e)If cleaning fails to reduce the indication to acceptable amplitude (below 3.0 major divisions), document the indications on the inspection data sheet. Refer to Figure 210.
(f)Reposition the eddy current probe over the suspect area of the seal tooth and scan again several times to make sure that the previously observed indication can be repeated. If the calibration is correct and the indication is repeatable, the indication must be considered valid.
(g)If the indication is not repeatable or less than (3.0 major divisions), the indication is not valid and must be considered acceptable.
Subtask 72-00-31-280-001
F.Disposition criteria.
(1)Any seal tooth indications that are equal or exceed 3 major divisions, must be considered rejected by this procedure.
(2)If there is no obvious reason for a rejectable eddy current indication, mark and record the location of the seal tooth indication on Figure 210 and continue to process the part to perform the FPI of the seal teeth by the engine manual. FPI must be the only inspection used to determine the serviceability of the part once the coating has been removed.
(3)If the rejectable eddy current indication appears to be from visual condition (damaged tooth), mark and record the location of the seal tooth indication on Figure 210 and continue to process the part to perform FPI of the seal teeth by the engine manual. FPI must be the only inspection used to determine the serviceability of the part once the coating has been removed.
Subtask 72-00-31-280-002
G.As a minimum, record all requested information on the inspection data sheet and forward to the local GE Aviation Product Support Engineering representative.
* * * FOR ALL
Figure 201   HPTR CDP Seal - Coated Seal Teeth Inspection
* * * FOR ALL
Figure 202   Calibration Standard P/N: SGP-112
* * * FOR ALL
Figure 203   Calibration Standard Holder P/N: SGP-113
* * * FOR ALL
Figure 204   Probe P/N: US-2056
* * * FOR ALL
Figure 205   Probe Fixture P/N: JO57138
* * * FOR ALL
Figure 206   Calibration Setup
* * * FOR ALL
Figure 207   Inspection Setup
* * * FOR ALL
Figure 208   Calibration Display
* * * FOR ALL
Figure 209   Evaluation/Reject Display
* * * FOR ALL
Figure 210   Form 2388-1 - INSPECTION DATA SHEET